EVALUASI HASIL KINERJA ALAT KIMIA KLINIK ReiGed RD-240 DAN MINDRAY BS-120 PADA PEMERIKSAAN TRIGLISERIDA (METODE GPO-PAP)
ABSTRACT
EVALUATION OF THE PERFORMANCE RESULTS OF CLINICAL CHEMICAL
TOOLS ReiGED RD-240 AND MINDRAY BS-120 ON TRIGLYCERIDE
EXAMINATION (GPO-PAP METHOD )
Maswanti NS, Kafesa A, Fadhila F.
The operating laboratory must be guaranteed its quality and validity, so
all laboratory operational methods and procedures must be integrated from
planning, taking test samples, handling, testing to providing reports on laboratory
test results to customers.
This study was conducted to determine the results of the evaluation of
the performance of the ReiGed RD-240 and Mindray BS-120 clinical chemistry
tools on triglyceride examination.
This research method is an analytic study with a Comparative Study
approach, namely by comparing the performance of two tools, namely the
Mindray BS-120 and ReiGed RD-240 tools.
The results obtained 2 times the +3SD results on the ReiGed RD-240
device while the Mindray BS-120 had the same results for 4 days, sigma 5 on
Reiged which means Excellent and sigma 6 on mindray, namely Word class. In the
normality test p = 0.000 (